Executive Summary



This Alert is flagged as TOP 25 Common Weakness Enumeration from CWE/SANS. For more information, you can read this.
Informations
Name CVE-2020-6963 First vendor Publication 2020-01-24
Vendor Cve Last vendor Modification 2024-11-21

Security-Database Scoring CVSS v3

Cvss vector : CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Overall CVSS Score 10
Base Score 10 Environmental Score 10
impact SubScore 6 Temporal Score 10
Exploitabality Sub Score 3.9
 
Attack Vector Network Attack Complexity Low
Privileges Required None User Interaction None
Scope Changed Confidentiality Impact High
Integrity Impact High Availability Impact High
Calculate full CVSS 3.0 Vectors scores

Security-Database Scoring CVSS v2

Cvss vector : (AV:N/AC:L/Au:N/C:C/I:C/A:C)
Cvss Base Score 10 Attack Range Network
Cvss Impact Score 10 Attack Complexity Low
Cvss Expoit Score 10 Authentication None Required
Calculate full CVSS 2.0 Vectors scores

Detail

In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilized hard coded SMB credentials, which may allow an attacker to remotely execute arbitrary code.

Original Source

Url : http://cve.mitre.org/cgi-bin/cvename.cgi?name=CVE-2020-6963

CWE : Common Weakness Enumeration

% Id Name
50 % CWE-798 Use of Hard-coded Credentials (CWE/SANS Top 25)
50 % CWE-20 Improper Input Validation

CPE : Common Platform Enumeration

TypeDescriptionCount
Os 1
Os 1
Os 2
Os 2

Sources (Detail)

https://www.us-cert.gov/ics/advisories/icsma-20-023-01
https://www3.gehealthcare.com/~/media/downloads/us/support/site-planning/site...
Source Url

Alert History

If you want to see full details history, please login or register.
0
1
Date Informations
2024-11-28 13:47:51
  • Multiple Updates
2020-05-23 02:40:28
  • First insertion